NV-1-488 / iFLiP Key Specs

Summary of core hardware, detection, timing, and software capabilities for the NV-1-488 / iFLiP fluorescence lifetime fiber photometry system.

Operation conditions

  • Operation voltage: 110 or 220 V, preset
  • Operation temperature: 18 – 25 °C (65 – 77 °F)
  • Operation humidity: ≤ 70%

Light source

  • Wavelength: 488 ± 10 nm
  • Pulse frequency: 80 MHz
  • Pulse width: ≤ 120 ps
  • Power output: ≥ 50 µW (typically ≥ 100 µW)
  • External intensity modulation: 0 – 5 V (5 V = minimum)
  • Max modulation frequency: 1 kHz
  • Light attenuation: > 20 dB (typically 25–30 dB)
  • Pulse sync signal: ≥ 400 mV

Signal detector

  • Single photon signal amplitude: ≥ 50 mV
  • Single photon signal FWHM: 4 ns
  • Quantum yield: 40%
  • Dark count: ≤ 25 kCPS (typically 15 kCPS)

TCSPC lifetime measurement module

  • Max photon count: 40 MCPS
  • Time resolution: ≥ 25 ps
  • Dead time: 25 ns
  • Single photon detection: CFD

Software

  • Live and post-hoc intensity display: Yes
  • Live and post-hoc MPET display: Yes
  • Live lifetime curve display: Yes
  • Post-hoc lifetime curve fitting: Yes
  • Live phasor transformation display: Yes
  • Optional background subtraction: Yes
  • Looping (adjustable fixed interval): Yes

Integrated system

  • Number of channels: 1
  • Sampling frequency: 0 – 5,000 Hz
  • Continuous acquisition duration: 0 – 100 hr (limited by hard drive space and other computer physical parameters)
  • System impulse response function: σ < 200 ps (of an approximating Gaussian)
  • External trigger: Supported
  • Recording external TTL sync signal: Supported
  • Acquisition status output: TTL on during acquisition